6KV EHT Unit HV3
This Instrument is now obsolete.
Dual 2KV EHT Unit HV2
This power supply provides two independent channels of positive polarity supply from 0 to 2kV at a maximum current of 200uA.
Each channel has a 3.5 digit DPM which can be switched to indicate output voltage, output current/ or current trip set point. Full scale ranges of 200V and 2kV, 20uA and 200uA are provided. It is housed in a double width NIM unit and draws current from the +24V, +12V and –12V NIM supplies.
This EHT supply is ideal for polarising semiconductor detectors.
External control is possible via a 25 way 'D' connector.
Specifications:
| Output Connector
| SHV
|
| Output Voltage
| 0 to 2kV
|
| Output Current
| 200uA max
|
| Load Regulation
| 0.01% (0 to 200uA)
|
| Temperature Stability
| 0.01%/°C max
|
| Noise & Ripple
| 10mVp-p max
|
| Trip Set
| 0 to 200uA
|
| Supply Requirements
| +24V @ 165mA
|
|
| +12V @ 310mA
|
|
| -12V @ 180mA
|