The CGM5 Radiation Monitor is a wall mounted self contained gamma radiation monitoring instrument, with gamma flux measurement range of 0.1mRad/Hr to 1000mRad/Hr. It is intended for use as an area gamma radiation alarm.
Indication is via a front panel mounted meter with a logarithmic characteristic. Comprehensive trip facilities and an audible high flux alarm are provided, together with remote indication ability.
The detector, an argon filled ionization chamber, is mounted integral with the head amplifier which may be operated remote from the Monitor. Remote indication of flux level is normally via 1mA moving coil meters.
An optional internally mounted eight bit ADC is available for digital data collection systems.
Specifications:
| Measuring Range
| 0.1 mRad/hr to 1000 mRad/hr
|
| Response Time Constant
| <2 seconds
|
| Accuracy
| ±10%
|
| Fault Trip
| Indicates equipment fault.
|
|
| Indicator light and volt-free output contacts.
|
| High Trip 1
| Indicates flux level is above High Trip Level 1.
|
|
| Amber indicator.
|
| High Trip 2
| Indicates flux level is above High Trip Level 2.
|
|
| Red indicator and volt-free output contacts.
|
| Ouput Contact Ratings
| 50V, 1A
|
| Trip Contacts Connector
| 6 pole Plessey Mk 4
|
| Remote Interface
| Two 1mA FSD flux signal outputs
|
|
| Remote 'recover'
|
|
| Remote EHT on/off
|
|
| Remote indicator lights
|
| Remote Connector
| 12 pole Plessey Mk 4
|
| Digital Output Option
| Provides an 8 bit output from an optional internal ADC
|
| Digital Output Connector
| 12 Pole Plessey Mk 4
|
| Head Connector
| 8 pole size 2 Lemo
|
| Test Input
| BNC on head allows injection of a dc test current into the ion chamber amplifier. Sensitivity is 5x10-13A/mRad/hr
|
| Mains Supply
| 240V 50Hz ±10% 15VA
|
| Main Unit Dimensions
| 405 x 240 x 153mm deep
|
| Head Dimensions
| 157 x 290 x 103mm deep
|
| Main Unit Weight
| 7kg
|
| Head Weight
| 3kg
|
Front panel monitor points of internal supply and signal voltages are availble on the front panel for test purposes.